Advocating for the implementation of SonoVue microbubbles as pressure sensors: a call to action for clinical noninvasive pressure estimation
Editorial Commentary

Advocating for the implementation of SonoVue microbubbles as pressure sensors: a call to action for clinical noninvasive pressure estimation

Ao Wen1,2,3#, Lingxiao Yang1,2,3#, Tao Lv1,2, Huayu Yang4, Fei Li5, Yilei Mao4, Gang Xu1,2, Jia-Yin Yang1,2

1Liver Transplant Center, Organ Transplant Center, West China Hospital of Sichuan University, Chengdu, China; 2Laboratory of Liver Transplantation, Key Laboratory of Transplant Engineering and Immunology, National Health Commission (NHC), West China Hospital of Sichuan University, Chengdu, China; 3West China Hospital, West China School of Medicine, Sichuan University, Chengdu, China; 4Department of Liver Surgery, Peking Union Medical College (PUMC) Hospital, PUMC and Chinese Academy of Medical Sciences, Beijing, China; 5Paul C. Lauterbur Research Center for Biomedical Imaging, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China

#These authors contributed equally to this work.

Correspondence to: Jia-Yin Yang, MD, PhD; Gang Xu, MD, PhD. Liver Transplant Center, Organ Transplant Center, West China Hospital of Sichuan University, Chengdu, China; Laboratory of Liver Transplantation, Key Laboratory of Transplant Engineering and Immunology, National Health Commission (NHC), West China Hospital of Sichuan University, 37 Guoxue Lane, Wuhou District, Chengdu 610041, China. Email: doctoryjy@scu.edu.cn; gangxu@wchscu.cn.

Keywords: SonoVue microbubbles; pressure estimation; noninvasive


Submitted Apr 07, 2024. Accepted for publication May 20, 2024. Published online Jul 18, 2024.

doi: 10.21037/hbsn-24-193


Pressure measurement within the body is of pivotal significance in the diagnosis of vascular and organ-related diseases associated with hydrostatic pressure. At present, the most commonly used clinical method is to insert a catheter along with a pressure sensor and then guide it to the area of interest through vessels such as central venous pressure (CVP). However, the presence of sensors within the vessel of interest will inevitably cause alterations to the circulation and thus affect blood pressure. Moreover, the use of invasive methods does not allow monitoring of every area inside the body.

Ultrasound contrast agents (UCAs), which are microbubbles, comprise a core of gas encased within a membrane made of lipids, proteins, or polymers. These agents are employed in contrast-enhanced ultrasonography (CEUS). UCA microbubbles can induce nonlinear oscillatory behaviors in the acoustic pressure field generated by ultrasound pulses (1). The nonlinear scattering signals of UCAs span a broad spectrum of frequencies, encompassing subharmonic frequencies (1/2f). And the subharmonic amplitude of the subharmonic scattering signal at a frequency 1/2ftransmit shows better ambient pressure sensitivity compared to fundamental harmonic amplitude (2). Additionally, these signals can be distinguished from those of tissue, improving the signal-to-noise ratio and facilitating the development of contrast-specific imaging modalities (3). Given that the acoustic properties of these gas-filled UCA microbubbles are closely correlated with the ambient pressure, they can serve as pressure sensors upon exposure to appropriate ultrasound stimulation (4). Allowing for a clinically acceptable level of sensitivity, the variation in subharmonic amplitude of the echo scattered by microbubbles has emerged as the most promising and suitable approach for clinical application. Presently, various studies have established a linear relationship between ambient pressure and the subharmonic signaling components emitted by various UCAs (5-7).

The team, directed by Prof. Forsberg, pioneered the revelation that UCAs could serve as noninvasive tools for pressure measurement, functioning as pressure sensors, through the observation of variations in the amplitude of subharmonic contrast bubble signals (2). This groundbreaking and innovative approach, termed subharmonic-aided pressure estimation (SHAPE), is predicated on the inverse linear relationship (r2>0.90) between the amplitude of subharmonic signals and the ambient hydrostatic pressure. It is advocated for noninvasive pressure determination (2,8,9). The emergence of SHAPE enables the noninvasive measurement of absolute pressures and pressure gradients throughout the body. To date, extensive clinical trials have concentrated on the utilization of SHAPE in conjunction with Sonazoid for the diagnosis of clinically significant portal hypertension (PH) (9,10). PH represents a significant complication of cirrhosis and is deemed pivotal in the success of liver transplantation. However, validated non-invasive techniques for its measurement are currently lacking (11,12). In a 2013 study, Eisenbrey et al. (9) examined the relationship between SHAPE data and the hepatic venous pressure gradient (HVPG). The gradient derived from the SHAPE from the portal to hepatic veins demonstrated a strong concordance with the HVPG (R=0.82). Patients identified as having an elevated risk for variceal hemorrhage (HVPG ≥12 mmHg) had a significantly greater mean subharmonic gradient than patients with lower HVPGs (1.93±0.61 vs. −1.47±0.29 dB, P<0.001), accompanied by 100% sensitivity and 81% specificity, suggesting the potential of the SHAPE as an effective screening instrument for the diagnosis and prediction of PH in patients undergoing trans-jugular liver biopsy. A prospective, multicenter trial conducted by Gupta et al. (10) reported that participants diagnosed with PH (n=21), characterized by an HVPG of 10 mmHg or greater, exhibited a greater SHAPE gradient than those with lower HVPGs (n=91) (0.27±2.13 vs. −5.34±3.29 dB; P<0.001), and the SHAPE achieved a sensitivity of 91% [95% confidence interval (CI): 88–93%] and a specificity of 82% (95% CI: 75–85%), corroborating that the SHAPE correlates well with the HVPG and allows the identification of participants with PH and those at elevated risk of variceal bleeding. In addition, despite limited sample sizes, SHAPE and three-dimensional subharmonic imaging have demonstrated their potential for estimating imaging responses and outcomes to neoadjuvant chemotherapy in patients with breast cancer as early as after the completion of 10% of neoadjuvant therapy (13).

To date, some research has also made essential contributions to the application of SHAPE under other ambient pressures. Esposito et al. (14) extended the SHAPE with Definity microbubbles to detect intracardiac pressures noninvasively and accurately and found that the root mean square, mean, and median errors across data acquired from the right ventricle and left ventricle were <5 mmHg. Kalayeh et al. (15) demonstrated the applicability of the SHAPE in measuring bladder pressure based on the strong relationship between the SHAPE data and bladder phantom pressure. Li et al. (16) reported that the SHAPE technique can estimate alterations in pressure within the vascular system and has the potential to introduce a novel approach for monitoring the biomechanical properties of the carotid artery, as well as evaluating the vulnerability of carotid plaque. Qiao et al. (17) utilized SHAPE in conjunction with plane wave transmission to noninvasively assess the pressure distribution and fractional flow within the middle cerebral artery in an in vitro setting and found that the fractional flow value was highly consistent with the value measured by a sensor.

In the second generation of UCAs, both SonoVue and Sonazoid microbubbles exhibit greater stability but have structural differences from each other (18). As shown in Table 1, Sonazoid is characterized by hydrogenated egg phosphatidyl serine and a perfluorobutane core, whereas SonoVue consists of a single phospholipid shell encapsulating an inert sulfur hexafluoride gas core. The above factors determine the potential difference in functional properties between the two in SHAPE. The function of SonoVue, which is mainly marketed and widely used in clinical diagnostics in China, as a pressure sensor in SHAPE technology has been less studied, and the conclusion is still unclear (7,19,20). Consequently, a series of experiments were undertaken by our research group with commercially accessible SonoVue microbubbles. These experiments explored the effect of acoustic parameters on the sensitivity of these microbubbles to pressure, as well as their potential for noninvasive pressure assessment. The outcomes garnered from these investigations were substantial and noteworthy. A previous experiment was carried out at ambient pressures in the range of 8–180 mmHg with an average gradient of change of 43 mmHg (19). The results indicated that SonoVue microbubbles under medium mechanical index (0.25–0.4) excitation exhibited high sensitivity and linearity for ambient pressure estimation based on subharmonic amplitude. Our in vitro experimental study carried out in a closed-loop dynamic flow experimental system demonstrated that the subharmonic amplitude of SonoVue microbubbles occurred in three stages with different incident acoustic pressures at 10 and 40 mmHg ambient pressure: the first growth stage (40–300 kPa), the saturation stage (300–400 kPa), and the second growth stage (400–540 kPa). In the first growth stage, the subharmonic amplitude increased with increasing ambient pressure (17). In the second growth stage, the subharmonic amplitude decreases with increasing ambient pressure and has the best correlation at an incident sound pressure of 520 kPa (r2=0.99), possibly originating from microbubble destruction during the second growth stage (21). Similar experiments by Nio et al. reported that the subharmonic amplitude of SonoVue microbubbles increased as the ambient pressure increased. In their study, within the acoustic pressure range of 20–500 kPa at a frequency of 4 MHz, the subharmonic amplitude of SonoVue microbubbles increased with increasing ambient pressure, reaching its peak at 50 mmHg (7). However, they only observed one growth phase, which is different from our study. In the second growth stage, we observed that the amplitude of subharmonics decreased significantly when the sound pressure was between 450 and 500 kPa (19). These results are consistent with the subharmonic amplitude increase in the sound pressure curve of another experimental phospholipid-coated microbubble (Bracco Research S.A., Geneva, Switzerland) explored by Frinking et al. (22). Based on the conclusion that the subharmonic amplitude of the second growth phase of SonoVue correlates more strongly with ambient pressure and root mean square error than that of the first growth phase, Lu et al. established animal models of PH to carry out in vivo experiments (23). The findings revealed a linear association between the subharmonic amplitude of SonoVue microbubbles and portal venous pressure (PVP) (<20 mmHg) at an incident sound pressure of 453 kPa. Subsequent validation of this discovery was conducted through in vivo experimentation utilizing a canine model of pulmonary hypertension by Xu et al. (24). When the incident sound pressure was set to 563 kPa, the diagnostic performance for PH was 93.3%, the specificity was 91.7%, and the overall accuracy was 92.6%. These experimental results show that the subharmonic scattering of SonoVue provides a promising noninvasive method for evaluating PVP.

Table 1

Differences between Sonazoid and SonoVue contrast microbubbles

Contrast agent microbubbles Physical structure and acoustic properties
Shell material Core gas Average diameter (μm) Duration in vivo (min) Frequency of resonance (MHz)
SonoVueTM Phospholipid Sulfur hexafluoride 2.6 5 2.0
SonazoidTM Hydrogenated egg phosphatidyl serine Perfluorobutane 2. 1 30 4.4

Currently, numerous clinical trials involving diverse ambient pressures and types of UCAs are underway investigating various applications of SHAPE in Asia, Europe, and North America (Table 2). Building upon the substantial evidence provided by existing studies (21,23,24), which suggest that subharmonic scattering signals derived from SonoVue microbubbles are a viable noninvasive alternative for assessing PVP in vivo, forthcoming research endeavors aim to evaluate the practical applicability of this methodology in a clinical setting. Moreover, the subharmonic scattering signal from SonoVue microbubbles has the highest sensitivity, accuracy, and specificity under specific conditions and can be extended to pressure measurements at other ambient pressures, such as ventricular pressure, carotid artery pressure, and bladder pressure. Although the results of the clinical trials are still pending, this nevertheless suggests that we may soon witness the integration of SHAPE into standard clinical practice.

Table 2

Clinical trials investigating various applications of SHAPE with diverse ambient pressures and types of UCA

Manometry environment Type of
UCA
Status Phase Country Interventional Model Study start date ID
Portal vein pressures Sonazoid Completed Phase 4 US Single group assignment 2015-04-06 NCT02489045
Sonazoid/Definity Recruiting Phase 3 US/Switzerland Parallel assignment 2022-11-28 NCT05470205
Sonazoid/Lumison (SonoVue) Recruiting Phase 2 US Parallel assignment 2021-11-03 NCT04720456
Intracardiac pressure Sonazoid Completed Phase 2 US Single center study/single group assignment 2017-12-04 NCT03245255
Definity Completed Phase 2 US Single group assignment 2017-06-01 NCT03243942
Pressure in the cancer Definity Recruiting Phase 2 US Single group assignment 2020-12-07 NCT04715958
Phase 3
Definity Recruiting Early phase 2 US Single group assignment 2020-11-24 NCT04721886
Unknown unknown Phase 0 China Unknown 2020-06-01 ChiCTR2000032911
Pressure gradient across the carotid plaque cap Definity Active, not recruiting Phase 4 US Single group assignment 2020-12-01 NCT04643431
Sonazoid Completed Unknown China Unknown 2021-05 ChiCTR1900027295
Others Definity Completed Phase 2 US Parallel assignment 2020-07-01 NCT05247541
Phase 3

SHAPE, subharmonic-aided pressure estimation; UCA, ultrasound contrast agent; US, United States.


Acknowledgments

Funding: This study was funded by the National Natural Science Foundation of China (No. 82300663), the Natural Science Foundation of Sichuan Province (No. 2022NSFSC0843), the Fundamental Research Funds for the Central Universities (No. 2023SCU12055) and the China Postdoctoral Science Foundation (No. 2022M712262).


Footnote

Provenance and Peer Review: This article was commissioned by the editorial office, HepatoBiliary Surgery and Nutrition. The article did not undergo external peer review.

Conflicts of Interest: All authors have completed the ICMJE uniform disclosure form (available at https://hbsn.amegroups.com/article/view/10.21037/hbsn-24-193/coif). H.Y. serves as an unpaid editorial board member of HepatoBiliary Surgery and Nutrition. Y.M. serves as the Editor-in-Chief of HepatoBiliary Surgery and Nutrition. The other authors have no conflicts of interest to declare.

Ethical Statement: The authors are accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. Written informed consent was obtained from the patient for publication of this manuscript.

Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/.


References

  1. Overvelde M, Garbin V, Sijl J, et al. Nonlinear shell behavior of phospholipid-coated microbubbles. Ultrasound Med Biol 2010;36:2080-92. [Crossref] [PubMed]
  2. Shi WT, Forsberg F, Raichlen JS, et al. Pressure dependence of subharmonic signals from contrast microbubbles. Ultrasound Med Biol 1999;25:275-83. [Crossref] [PubMed]
  3. Chong WK, Papadopoulou V, Dayton PA. Imaging with ultrasound contrast agents: current status and future. Abdom Radiol (NY) 2018;43:762-72. [Crossref] [PubMed]
  4. Pang M, Yang H. Subharmonic scattering of ultrasound contrast agent microbubbles may be an effective and promising tool for portal vein pressure estimation. Hepatobiliary Surg Nutr 2023;12:625-7. [Crossref] [PubMed]
  5. Halldorsdottir VG, Dave JK, Leodore LM, et al. Subharmonic contrast microbubble signals for noninvasive pressure estimation under static and dynamic flow conditions. Ultrason Imaging 2011;33:153-64. [Crossref] [PubMed]
  6. Liu S, Wu J, Gu Y, et al. Ambient Pressure Evaluation Through Sub-Harmonic Response of Chirp-Sonicated Microbubbles. Ultrasound Med Biol 2017;43:332-40. [Crossref] [PubMed]
  7. Nio AQX, Faraci A, Christensen-Jeffries K, et al. Optimal Control of SonoVue Microbubbles to Estimate Hydrostatic Pressure. IEEE Trans Ultrason Ferroelectr Freq Control 2020;67:557-67. [Crossref] [PubMed]
  8. Dave JK, Halldorsdottir VG, Eisenbrey JR, et al. On the implementation of an automated acoustic output optimization algorithm for subharmonic aided pressure estimation. Ultrasonics 2013;53:880-8. [Crossref] [PubMed]
  9. Eisenbrey JR, Dave JK, Halldorsdottir VG, et al. Chronic liver disease: noninvasive subharmonic aided pressure estimation of hepatic venous pressure gradient. Radiology 2013;268:581-8. [Crossref] [PubMed]
  10. Gupta I, Eisenbrey JR, Machado P, et al. Diagnosing Portal Hypertension with Noninvasive Subharmonic Pressure Estimates from a US Contrast Agent. Radiology 2021;298:104-11. [Crossref] [PubMed]
  11. Eshkenazy R, Dreznik Y, Lahat E, et al. Small for size liver remnant following resection: prevention and management. Hepatobiliary Surg Nutr 2014;3:303-12. [PubMed]
  12. Jeong WK. Managing portal hypertension in patients with liver cirrhosis and hepatocellular carcinoma: non-invasive diagnosis and systemic treatment considerations. Hepatobiliary Surg Nutr 2023;12:960-2. [Crossref] [PubMed]
  13. Nam K, Eisenbrey JR, Stanczak M, et al. Monitoring Neoadjuvant Chemotherapy for Breast Cancer by Using Three-dimensional Subharmonic Aided Pressure Estimation and Imaging with US Contrast Agents: Preliminary Experience. Radiology 2017;285:53-62. [Crossref] [PubMed]
  14. Esposito C, Machado P, McDonald ME, et al. Noninvasive Evaluation of Cardiac Chamber Pressures Using Subharmonic-Aided Pressure Estimation With Definity Microbubbles. JACC Cardiovasc Imaging 2023;16:224-35. [Crossref] [PubMed]
  15. Kalayeh K, Fowlkes JB, Chen A, et al. Pressure Measurement in a Bladder Phantom Using Contrast-Enhanced Ultrasonography-A Path to a Catheter-Free Voiding Cystometrogram. Invest Radiol 2023;58:181-9. [Crossref] [PubMed]
  16. Li R, Zhang Y, Zheng S, et al. Biomechanical characteristics in the carotid artery: Noninvasive assessment using subharmonic emissions from microbubbles. Med Phys 2023;50:6857-63. [Crossref] [PubMed]
  17. Qiao X, Zhang R, Yu J, et al. Noninvasive assessment of pressure distribution and fractional flow in middle cerebral artery using microbubbles and plane wave in vitro. Ultrasonics 2024;138:107244. [Crossref] [PubMed]
  18. Claudon M, Dietrich CF, Choi BI, et al. Guidelines and good clinical practice recommendations for contrast enhanced ultrasound (CEUS) in the liver--update 2012: a WFUMB-EFSUMB initiative in cooperation with representatives of AFSUMB, AIUM, ASUM, FLAUS and ICUS. Ultraschall Med 2013;34:11-29. [PubMed]
  19. Li F, Li D, Yan F. Improvement of Detection Sensitivity of Microbubbles as Sensors to Detect Ambient Pressure. Sensors (Basel) 2018;18:4083. [Crossref] [PubMed]
  20. Andersen KS, Jensen JA. Impact of acoustic pressure on ambient pressure estimation using ultrasound contrast agent. Ultrasonics 2010;50:294-9. [Crossref] [PubMed]
  21. Xu G, Lu H, Yang H, et al. Subharmonic Scattering of SonoVue Microbubbles Within 10-40-mmHg Overpressures In Vitro. IEEE Trans Ultrason Ferroelectr Freq Control 2021;68:3583-91. [Crossref] [PubMed]
  22. Frinking PJ, Brochot J, Arditi M. Subharmonic scattering of phospholipid-shell microbubbles at low acoustic pressure amplitudes. IEEE Trans Ultrason Ferroelectr Freq Control 2010;57:1762-71. [Crossref] [PubMed]
  23. Lu H, Xu G, Wang Y, et al. Correlation Between Portal Vein Pressure and Subharmonic Scattering Signals From SonoVue Microbubbles in Canines. Ultrasound Med Biol 2023;49:203-11. [Crossref] [PubMed]
  24. Xu G, Wang Y, Lu H, et al. Portal Vein Pressure Estimation and Portal Hypertension Discrimination Based on Subharmonic Scattering of Ultrasound Contrast Agent Microbubbles. IEEE Trans Biomed Eng 2024;71:150-9. [Crossref] [PubMed]
Cite this article as: Wen A, Yang L, Lv T, Yang H, Li F, Mao Y, Xu G, Yang JY. Advocating for the implementation of SonoVue microbubbles as pressure sensors: a call to action for clinical noninvasive pressure estimation. HepatoBiliary Surg Nutr 2024;13(4):690-695. doi: 10.21037/hbsn-24-193

Download Citation